medical laser optical fiber
accessory applied with medical laser therapeutic apparantus, transmiting energy generated by laser, can be used in Laser surge
Performance Structure And Composition
Composed by optical fiber connector protective cap
specification | (μm) optical fiber dia. | |
Disposable sterile medical Laser optical fiber | YYGX200,YYGX2001,YYGX2002,YYGX20012 | 200μm±10% |
YYGX400,YYGX4001,YYGX4002,YYGX40012 | 400μm±10% | |
YYGX600,YYGX6001,YYGX6002,YYGX60012 | 600μm±10% | |
Non-sterile medical Laser optical fiber | YYGX200F,YYGX2001F,YYGX2002F,YYGX20012F | 200μm±10% |
YYGX400F,YYGX4001F,YYGX4002F,YYGX40012F | 400μm±10% | |
YYGX600F,YYGX6001F,YYGX6002F,YYGX60012F | 600μm±10% |
1.Appearance: fiber surface should be clean,seamless and burrfree
2.Transmission efficiency: should be greater than 70%
3.folding strength
(um) Fiber dia. | 200 | 400 | 600 |
(mm) Folding dia. | ≥25 | ≥35 | ≥45 |
biological property of Medical laser optical fiber
4.axenic : (One-off axenic medical Laser optical fiber)
5.Cytotoxicity: not greater than level I
6.intracutaneous stimulation: not greater than slight stimulation
7.delayed type hypersensitivity: should be no this hypersensitivity
8.Hemolysis: not greater than 5.0%
9.systemic toxicity test: should be no this toxicity
Chemical property of Medical laser optical fiber
10.Reducing substances :potassium permanganate consuming volume difference between test liquid and comparing liquid Should not be greater than 2ml
11.pH value: pH value difference between test liquid and comparing liquid Should not be greater than 1
12.heavy metal concentration: total heavy metal concentration in test liquid Should not be greater than 5μg/ml,cadmium concentration not be greater than 0.1μg/ml
13.Epoxy Ethane Residue: should not be greater than 10μg/g in optical fiber Sterilized by Epoxy Ethane (only apply to Disposable sterile medical Laser optical fiber)